Collecting Very Fine Volumetric Accuracy Data

As part of the V&V (validation and verification) test procedure we developed a while ago, we designed a gravimetric means to characterize the volumetric and flow-rate accuracy of a couple of precision injection instruments (which I thought I’d share with you today). With this setup, some programming, and a fairly precise scale, we can automatically collect data during an injection to determine if the device meets the accuracy tolerances from the specification. We’re able to achieve measurement resolution of just 0.1 milligram and reduced the error due to water evaporation to 40 times better than using a mineral oil emulsion.

The Scale

We chose an Acculab 210.4 scale because it was accurate enough (.0001 gram resolution, which is about a tenth of a microliter for room temperature water) and because it has a serial interface that we can call from Visual Basic to easily collect data and put into a spreadsheet. The data stream is limited to about 8Hz, but it was fast enough for our purposes.

Scales with more resolution often come with damping mechanisms to keep them stable. Although the glass cover keeps the air movement down, this one didn’t have a means to damp vibrations, so we needed to isolate the scale ourselves. We used a small table with some rubber feet as the foundation. Then we used one of those heavy engineering textbooks we have so many of on some thick foam and put the scale on top. Considerate use of flexible beams, soft materials, and masses meant the scale was rock-steady without any software averaging required.

The Fluid Connection

To get the fluid onto the scale, we attached tubing with a luer-lock fitting to a long needle inserted into the scale cover through a hole in the top plate and held in place with a thumb-screw. By using a needle instead of just inserting the tubing into the container, we eliminated any contact with the container or scale. we found that any such contact threw the scale out of whack, especially when the fluid flowed through the tubing with momentum.

The question came up as to whether the needle should be above the water-level or below it. If the needle is below the water-level, surface tension on the needle can reduce the weight of the water. However, if the needle is above the water-level, a drop can form at the tip instead of depositing the small volume onto the scale. We needed fine-resolution flow-rate data, so we opted to ignore the surface tension, which we found to be below the resolution of the scale.

Evaporation of the Water

Over the duration of a test, water will evaporate. The rate is dependent upon temperature, relative humidity, and the surface area of the container. Many suggest adding mineral oil to the water to create a barrier layer to supposedly eliminate evaporation of the water. Data showed that the oil layer did reduce the rate of evaporation (from 0.13 g/hr to 0.08 g/hr), although the oil ruined the plastic connections of the tubing by making them brittle and causing them to leak and break. Instead, we drilled a 1/4” hole into the cap of the container for the needle to fit through, which dropped the rate of evaporation by two orders of magnitude. We collected data for various configurations of the water, oil, and lid and plotted them below. So, without using any oil, we were able to drop the rate of evaporation from 0.127 g/hr to 0.002 g/hr.

Volume and Flow Rate

So, how does a measure of weight relate to volume and flow-rate? Well, the density of water is pretty well defined based on temperature (and weight / density = volume). Accounting for the error of this density value, as well as errors in the scale resolution and time resolution provided excellent measurement results by which to evaluate the devices.